Standard Features
- High impact sealed case provides improved component protection.
- Radiation detected Gamma or X-rays above 60 Kev.
- Response time 90% of final reading in 7 seconds.
- Low battery indicator.
- Detector energy compensated, halogen quenched, internally mounted Geiger tube.
- Precision within 10% at all levels.
Specific Unit Features |
|||
Feature |
Model 4 |
Model 2S |
Model 2D |
Analog Meter-rugged assembly with shatterproof lexan lens. |
No |
Yes |
Yes |
Range selection-multi-purpose range selection switch & battery test |
0 mR/hr through 2 R/hr |
0-10, 0-100, 0-1000 mR/hr |
0-10, 0-100, 0-1000 mR/hr |
Battery life: Hrs@ 4 hrs/dayDays constant use |
600 24 |
250 15 |
250 15 |
Illuminated digital display |
Yes |
No |
No |
Geotropism % of full scale |
Not Applicable |
±2 % |
±2 % |
Energy dependence (within 20%) |
80-1,000 Kev |
80-1,000 Kev |
80-1,000 Kev |
Battery Complement |
8 AA alkaline |
2 “D” Cell |
2 “D” Cell |
Audible alarm |
Automatic instantaneous over 250 mR/hr |
Optional |
Yes all ranges |
Calibration with Cs-137 |
± 10 % |
± 10 % |
± 10 % |
Temperature dependence |
± 5 %, -20°C to 55°C |
± 10 %, -30°C to 50°C |
± 10 %, -30°C to 50°C |
Weight |
2.9 lbs., 1.4 kg |
3 lbs., 1.5 kg |
3 lbs., 1.5 kg |
Dosimeter Charger & Audible alarm |
Yes |
No |
Yes |